SNASharp v2019_08_30_0

This new release of SNASharp embed some stuff usefulls for hardware analyzers conceptors.

The build can be directly downloaded here

There is not embeded installer, you must simply unpack the zip archive and run the executable.

Like previous releases you need .NET Framework 4.0 to run the program. Minimal supported operating system is Windows XP SP2, but it work fine on all recents versions of windows (32 and 64 bits). You can also run the program under Linux system (or Apple Mac OSX ) using Mono Framework

Under Mac OSX you need to force the 32 bits version of Mono ( .NET Winform is not fully supported on 64 bits architecture)

Command line : mono  — arch=32  SNASharp.exe

About this version..

I added the possibility to disable the calibration if needed. With this new capture mode you can evaluate the real bandwidth of your analyzer (without the calibration correction). This really help to evaluate an hardware improvment. To activate this mode you need to check « Raw capture« . The 0dB reference (for each analyzers) can be set on device editor panel on « RawMode_0dB_Reference » entry.


The displayed level is done by:

Displayed Level In dB = (Analyzer captured value – RawMode_0dB_Reference ) * VerticalResolutiondB .

Later don’t miss to re-enable the calibration to make accurate measurements (The raw mode capture option is automatically saved and restored when the program exit and start).

v2019_08_30_0 new options
D6 Analyzer capture in raw mode

Another new option is « Use Att cal » ( Checked by default). This option is dedicated for embeded attenuators analyzers ( like NTW70,NTW300,NWT500). If you keep this option checked, the behaviors stay the same as previous versions of SNASharp. If you unchek this option the analyzer will avoid to use specific attenuator levels calibration set ( In this case the 0dB calibration set serves as reference).

Why this option ?

It’s usefull to check if embeded attenuators of your analyzer works fine, and evaluate the linearity of your hardware. In addition, this can reassure some users who did not understand why the measurements remain on a reference of 0 dB while attenuators were engaged.

NWT70 capture with differents attenuator sets and « Use Att cal » unchecked

I added The Geekcreit® Spectrum Analyzer USB LTDZ_35-4400M on supported hardware list( not tested yet).

Geekcreit® Spectrum Analyzer USB LTDZ_35-4400M

Don’t hesitate to put a comment below this article if you have any question.